用于在线检测的紧凑型多测量模式干涉仪系统光路布局如图1所示,其中光源系统包含波长分别为632.8 nm和543 nm的气体激光器以及470 nm LED光源,并且三个波长光束可进行切换开启。来自光源系统的光经光纤耦合接入干涉系统的光纤准直器后变为准直光束,再经一个线偏振片得到某一方向的线偏振光。为了缩短系统尺寸,采用折光反射镜对光路进行90°转折。线偏振光分别经过分光棱镜(BS)和偏振分光棱镜(PBS)后分为透射p偏振光和反射s偏振光,其中p偏振光作为检测光,先后经长波通过滤光片(F1)(三波长光束均透射)和显微物镜1后入射到待测元件表面并沿原路反射回来;s偏振光作为参考光,其中632.8 nm和543 nm波长激光束被短波长通过滤光片(F2)第一个面反射回来,470 nm波长LED光束透射短波长通过滤光片(F2)后经显微物镜2入射到位于焦平面上的参考反射镜并沿原路返回。p偏振检测光和s偏振参考光沿原路返回并在偏振分光棱镜(PBS)透射、反射汇合后,分别经分光棱镜(BS)反射和宽波段消色差1/4波片(AQWP)透射后,变为两个旋向相反的圆偏振光。两束圆偏振光经成像透镜在偏振相机上可同时得到相移量为π/2的四个通道移相干涉条纹,利用四步移相算法即可计算得到与待测元件表面信息相对应的瞬态波前分布,进而实现待测元件表面的瞬态检测。
为了满足不同工作波长应用要求,系统中光纤准直器采用了离轴抛物线反射式光纤准直器,分光棱镜(BS)、偏振分光棱镜(PBS)和1/4波片(AQWP)都为宽工作波段。针对在线检测中环境扰动影响问题,系统采用了偏振相机来实现波前的四步偏振移相瞬态检测,以最大程度提高系统的抗干扰能力。偏振相机是在传统的象元面阵前加入了由2×2微偏振器(透光轴方向分别为0°、45°、90°和135°)为超像素单元所组成的偏振面阵(图1),因而可实现一帧图像同时获取四个偏振方向对应四个通道图像信息[12]。
-
为了利用单一光路来实现不同测量模式的切换,系统采用短波长通过滤光片(F2,截止波长:500 nm,反射波段:512~715 nm,透射波段:300~490 nm)实现激光光束的反射和LED光束的透射,同时利用长波通过滤光片(F1,截止波长:350 nm,反射波段:200~340 nm,透射波段:358~1 600 nm)来补偿光程。在单波长或多波长激光干涉测量模式中,可采用512~715 nm范围激光器作为光源,该模式下滤光片(F2)的第一入射面的反射光作为参考光;在干涉显微测量模式中,可采用358~490 nm中心波长范围的LED作为光源,该模式下参考反射镜的反射光作为参考光。
在单波长激光干涉测量模式或者LED显微测量模式中,利用偏振相机单帧采集得到四通道π/2移相条纹,设0°、45°、90°和135°透光轴方向对应的干涉图强度分布为
${I_j}$ (j=1, …, 4),可得对应的待测波前光程差为:$$ \Delta {\rm{ = }}\frac{{\lambda }}{2\pi } \cdot {\arctan}\left( {\frac{{{I_2} - {I_4}}}{{{I_1} - {I_3}}}} \right) $$ (1) 式中:
$\lambda $ 为光波长。在单波长干涉检测中,其表面相邻两个采样点可测高度差最大值为λ/4,即对应632.8 nm光波长为0.158 2 μm)。为了满足实际加工中到自由曲面等大动态范围面形测量需要,可采用多波长干涉技术来扩展干涉检测系统的测量范围。对于
${\lambda _1}$ 和${\lambda _2}$ 两个波长组合情况,其表面相邻两点可测高度差最大值为Λ/4,其中等效波长$\varLambda = {{{\lambda _1}{\lambda _2}}/{({\lambda _1} - {\lambda _2})}}$ 。对应两个波长为632.8 nm和543 nm的激光波长,其组合对应的等效波长为Λ=3.826 4 μm,若采用两个更为接近的波长则可得到更大的等效波长。 -
图1所示紧凑型多测量模式干涉仪中,主要误差因素包括滤光片面形误差、1/4波片相位延迟误差、偏振相机视场误差和多波长干涉误差等。
在宏观面形干涉检测中,滤光片(F1)的透射像差和滤光片(F2)的入射面(为参考面)会对检测结果引入误差;微观干涉显微测量中,滤光片(F1)和滤光片(F2)的透射像差会对检测结果引入误差。为了最大程度减小滤光片对测量结果的影响,系统中滤光片玻璃基底选用面形精度较高的平板玻璃。实际系统中滤光片面形误差均方根(RMS)约为10 nm量级,测量中可利用高精度干涉仪测得滤光片面形误差并作为系统误差,在最后被测结果中加以校正消除。
干涉系统利用1/4波片(AQWP)将振动方向相互垂直的两个线偏振测量光和参考光转换为旋向相反的圆偏振光。针对不同光波长应用要求,采用宽波段消色差1/4波片(AQWP),其对应470 、543 、和632.8 nm光波长的相位延迟量分别为0.251 7λ、0.245 4λ和0.246 0λ,而非理想情况下的λ/4。1/4波片对不同相位所引入的波前误差分布曲线如图2所示,对应470 、543 、632.8 nm光波长的波前测量误差峰谷(PV)值分别仅为0.9×10−5λ、6.6×10−5λ和5.0×10−5λ,因而可忽略其对测量结果的影响。
在对偏振相机所采集的单帧图像提取四个通道移相干涉条纹并根据式(1)进行相位或光程差求解时,近似认为同一超像素单元(对应2×2 pixel)内的相位分布保持不变,而该近似处理对测量结果会一定的测量误差,即偏振相机视场误差。偏振相机视场误差的大小与入射波前斜率有关,可利用相邻像素相位值的加权平均线性插值方法[13],对该误差因素实现有效的校正。图3所示为不同波前斜率对应偏振相机视场误差以及采用3×3相邻像元加权平均线性插值方法校正后引入的波前误差PV值。以图3所示的3×3相邻像元相位加权平均线性插值校正为例,各个通道的强度
${I_j}$ (j=1, …, 4,分别对应0°、45°、90°和135°透光轴方向)计算公式为:$$ \left\{ \begin{aligned} & {I_1} = {I_{m,n}} \\ & {I_2} = {{({I_{m - 1,n}} + {I_{m + 1,n}})}/2} \\ & {I_3} = {{({I_{m - 1,n - 1}} + {I_{m - 1,n + 1}} + {I_{m + 1,n - 1}} + {I_{m + 1,n + 1}})}/4} \\ & {I_4} = {{({I_{m,n - 1}} + {I_{m,n + 1}})}/2} \end{aligned} \right.$$ (2) 根据图3可知,偏振相机视场误差会随着波前斜率的增大而快速增大,对应0.02λ/pixel的波前斜率,其引入的波前误差PV值为0.251 7λ;利用相邻像元相位加权平均线性插值法可实现偏振相机视场误差的有效校正,校正后对应0.02λ/pixel的波前斜率引入的波前误差PV值仅约为5.0×10−5λ。
在多波长波前干涉检测中,通过等效波长概念可实现测量范围的有效放大,但其对应的相位噪声(如透射元件对不同波长的色散作用)也同时会按波长同等放大倍数被放大,进而会对最后测量结果引入误差,即多波长干涉误差。可采用条纹级次法[14-15]来校正多波长干涉误差,令两组
${\lambda _1}$ 和${\lambda _2}$ 不同光波长测得波面相位分别为${\varphi _1}$ 和${\varphi _2}$ ,则待测光程差$W$ 对应等效波长Λ的等效相位为$\psi = {\varphi _1} - {\varphi _2}$ :$$W = {{\varLambda \psi }/{2\pi }}$$ (3) 由此可将其转换为单波长
${\lambda _1}$ 对应测得光程差${W_1}:$ $$\left\{ \begin{aligned} & {W_1} = {\lambda _1}\left( {{{{\varphi _1}}/{2\pi }} + {N_1}} \right) \\ & {N_1} = {\rm{INT}}\left( {{W/{{\lambda _1}}}} \right) \end{aligned} \right.$$ (4) 式中:INT表示取整函数。当波长组数大于2时,可根据公式(4)类推。根据条纹级次法可以使多波长干涉测量获得与单波长测量相同的测量精度。
Compact multi-measurement mode interferometer for on-machine testing
-
摘要: 光学干涉仪由于高精度、全口径、非接触特点在光学元件的检测中具有极其广泛的应用。针对光学元件加工在线检测需要,提出了一种紧凑型的多测量模式瞬态干涉仪。该系统可同时实现单波长激光干涉、多波长激光干涉以及LED干涉显微测量等多种工作模式,以分别满足不同动态范围宏观面形以及表面粗糙度等显微结构的干涉检测。针对在线检测应用中复杂的环境振动影响,系统采用偏振相机来实现瞬态的偏振移相波前检测。为验证系统方案的可行性,对测量系统的主要误差因素进行了分析,并对不同工作模式下的金刚石车削机床在线检测结果与Zygo激光干涉仪和Zygo光学轮廓仪进行了比对实验,同时也利用多波长技术对自由曲面进行了在线检测。结果表明该系统可实现高精度的多测量模式,并且还可以满足大动态范围波前测量要求。该系统结构紧凑,整体尺寸仅为195 mm×160 mm×65 mm,极其适合车削机床的在线安装及检测。同时系统基于瞬态波前检测,具有对环境扰动不敏感的特点,在机床对中工具在线调整以及加工元件等在线检测中具有广泛的应用前景。Abstract: With the advantages of high precision, full field and non-contact testing, the interferometers have been widely applied in the testing of optical elements. To meet the need of on-machine testing of optical elements in the machining stage, a compact interferometer with multiple working modes was proposed to achieve transient testing. The proposed system can work in modes including single-wavelength laser interferometer, multi-wavelength interferometer and LED interferometric microscope, enabling the testing of surface figure with large dynamic range, surface roughness microscopic structures. A polarization camera was utilized to realize the simultaneous phase-shifting measurement with a single shot, by which the effect environmental disturbance can be minimized. To demonstrate the feasibility of the proposed interferometer, the on-machine testing in various working modes have been carried out on diamond turning machine, as well as the control experiment with Zygo laser interferometer and Zygo optical profiler; Besides, the freeform surface testing has also been achieved with the proposed interferometer. High measurement accuracy and large dynamic range has been achieved in experiments. It was compact in system with dimensions of 195 mm×160 mm×65 mm, making it extremely suitable for installing and testing on machining equipment. Besides, the transient wavefront testing makes the proposed system insensitive to environmental disturbance, and it has a great prospect for application in on-machine testing of alignment tool and various optical elements in machining stage.
-
Key words:
- on-machine testing /
- interferometer /
- transient wavefront /
- multiple working modes
-
[1] Nomura T, Yoshikawa K, Tashiro H, et al. On-machine shape measurement of workpiece surface with Fizeau interferometer [J]. Precision Engineering, 1992, 14(3): 155−159. doi: 10.1016/0141-6359(92)90004-G [2] Wang Daodang, Yang Yongying, Chen Chen, et al. Point diffraction interferometer with adjustable fringe contrast for testing spherical surfaces [J]. Applied Optics, 2011, 50(16): 2342−2348. doi: 10.1364/AO.50.002342 [3] Zhang Yu, Jin Chunshui, Ma Dongmei, et al. Key technology for fiber phase-shifting point diffraction interferometer [J]. Infrared and Laser Engineering, 2015, 44(1): 254−259. (in Chinese) [4] Wang D, Liang R. Simultaneous polarization Mirau interferometer based on pixelated polarization camera [J]. Optics Letters, 2016, 41(1): 41−44. doi: 10.1364/OL.41.000041 [5] Xie Fang, Wang Jian, Ma Sen, et al. Multiplexing optical fiber low coherence and high coherence interferometrc system with large range and high resolution for online measurement [J]. Measurement, 2013, 46(1): 171−176. doi: 10.1016/j.measurement.2012.06.004 [6] Zou Xicong, Zhao Xuesen, Li Guo, et al. Non-contact on-machine measurements using a chromatic confocal probe for an ultra-precision turning machine [J]. The International Journal of Advanced Manufacturing Technology, 2017, 90(5-8): 2163−2172. doi: 10.1007/s00170-016-9494-3 [7] Sims-Waterhouse D, Piano S, Leach R K. Verification of micro-scale photogrammetry for smooth three-dimensional object [J]. Measurement Science and Technology, 2017, 28(5): 055010. doi: 10.1088/1361-6501/aa6364 [8] Oh C J, Lowman A E, Smith G A, et al. Fabrication and testing of 4.2 m off-axis aspheric primary mirror of daniel K inouye solar telescope[C]// SPIE ,2016, 9912: 99120O. [9] Liu Dong, Yan Tianliang, Wang Daodang, et al. Review of fringe-projection profilometry and phase measuring deflectometry [J]. Infrared and Laser Engineering, 2017, 46(9): 0917001. (in Chinese) doi: 10.3788/IRLA201746.0917001 [10] Gong Zhidong, Wang Daodang, Wang Chao, et al. Misalignment calibration in reverse Hartmann surface test based on Zernike polynomials [J]. Chinese Journal of Scientific Instrument, 2018, 39(8): 178−184. (in Chinese) [11] Tian X, Zhang Y, Sohn A, et al. Dual-mode snapshot interferometric system for on-machine metrology [J]. Optical Engineering, 2019, 58(4): 044104. [12] Millerd J, Brock N, Hayes J, et al. Pixelated phase-mask dynamic interferometer[C]//SPIE, 2004, 5531: 304-314. [13] Novak M, Millerd J, Brock N, et al. Analysis of a micropolarizer array-based simultaneous phase-shifting interferometer [J]. Applied Optics, 2005, 44(32): 6861−6868. doi: 10.1364/AO.44.006861 [14] Creath K. Step height measurement using two-wavelength phase-shifting interferometry [J]. Applied Optics, 1987, 26(14): 2810−2816. doi: 10.1364/AO.26.002810 [15] Zhuo Yongmo, Li Tianping. The two-wavelength digital wavefront interferometry and the method for the error-reduction [J]. Journal of Zhejiang University (Natural Science), 1989, 4(23): 499−508. (in Chinese)