The DMD-based NIR spectrometer optical structure mainly includes a light source, a collimator lens, a grating, an imaging lens, a DMD, a focusing lens, a detector, a spectrum acquisition and processing system. The schematic structure is shown in Fig.1.
The light emitted from the light source passes through the slit and then enters the collimator lens. The collimator lens collimates the light into a multi-color parallel light through the sample cell, and the light that carries the sample information is incident on the planar reflective grating for diffraction and dispersion into a spectral band. Then, the image of the slit is imaged on the surface of the DMD by the imaging lens according to the wavelength. At this time, the DMD has two states: the incident light is reflected into the converging lens when the state is +12°, and the incident light is reflected into the absorbed by the absorption cell when the state is −12°. All the light entering the converging lens is converged on the detector for receiving. The detector measures the light intensity of different wavelengths in time, and the spectrum diagram containing the sample information is obtained after data processing.
-
The entrance slit is an important factor of the spectrometer. The smaller the slit is, the higher the resolution of the system is. On the contrary, the lower the resolution, the smaller the slit is, the higher the energy is in the system. According to the Rayleigh criterion, the resolution can be expressed as:
$$R = \frac{{\overline {\rm{\lambda }} }}{{\Delta \lambda }} = \frac{{\left( {{\lambda _1} + {\lambda _2}} \right)/2}}{{\left| {{\lambda _1} - {\lambda _2}} \right|}}$$ (1) Assuming that the entrance slit width is a, spectral broadening
$\Delta {\lambda }' $ can be expressed$ \dfrac{a}{{\rm{d}}l/{\rm{d}}\lambda } $ according to the grating dispersion formula, where$ {\rm{d}}l/{\rm{d}}\lambda $ is the liner dispersion of the grating[13-14]. Moving$ {\lambda }_{1} $ to the left$\Delta {\lambda }'/2 $ made the spectrum resolvable, but the resolution of the spectrum is reduced and the resolution expression is:$${{R}} = \frac{{{{\overline \lambda }^\prime }}}{{\Delta \lambda '}} = \frac{{({\lambda _1}^\prime + {\lambda _2})/2}}{{{\lambda _2} - {\lambda _1}^\prime }} = \frac{{4\overline \lambda + a/({\rm{d}}l/{\rm{d}}\lambda )}}{{4\Delta \lambda + 2a/({\rm{d}}l/{\rm{d}}\lambda )}}$$ (2) where
${\lambda _1}^\prime ={\lambda _1}+\Delta {\lambda ^\prime }/2$ -
The larger the focal length of the collimator lens is, the better the parallelism of the collimated light is. But when the system NA value is constant, increasing the focal length will increase the incident aperture, resulting in an increase in system aberration and reducing the spectral resolution; In addition, it also increases the system's size and the cost of the instrument. The focal length of the collimator lens can be chosen to balance the needs of the above two aspects. After the incident light is split by the grating, it is incident on the DMD, and only the micromirror with the DMD is at 12° to reflect the incident light to the subsequent optical system. According to the principle of geometrical optics, without considering slit diffraction and aberration, the linear dispersion of the DMD surface can be expressed:
$$\frac{{{\rm{d}}l}}{{{\rm{d}}\lambda }} = \frac{{\rm{{d}}\theta }}{{{\rm{d}}\lambda }}\frac{{{f_2}}}{{\cos \sigma }} = \frac{m}{{{{d}}\cos \theta }}\frac{{{f_2}}}{{\cos \sigma }}$$ (3) where σ is the angle between the central axis of the imaging lens and the focal plane normal. In the DMD operating state, σ is 12°.
$ {f}_{2} $ is the focal length of the imaging objective, the line dispersion rate is expressed as:$$\frac{{{\rm{d}}l}}{{{\rm{d}}\lambda }} = \frac{L}{{\Delta \lambda }}$$ (4) Using Eqs. (2), (3) and (4), the focal length of the imaging lens can be obtained:
$${f_2} = \frac{{{\rm{d}}l}}{{{\rm{d}}\lambda }} \times \frac{{d\cos \theta }}{m} \times \cos \sigma $$ (5) -
A lens is an optical element in which a transparent medium is interposed between two refractive surfaces. The refractive surface of the lens is mostly a spherical surface. According to the basic principles of geometric optics, the lens focal length is:
$$f'= - f = - \frac{{{f_1}^\prime {f_2}^\prime }}{\Delta } = \frac{{n{r_1}{r_2}}}{{(n - 1)[n({r_2} - {r_1}) + (n - 1)d]}}$$ (6) where
$ f $ is the focal length of the object,$ {f}^{\mathrm{\text{'}}} $ is the image focal length,$ {r}_{1} $ and$ {r}_{2} $ are the radius of curvature of the two spherical surfaces of the lens, n is the refractive index of the lens, and d is the thickness of the lens[15-16]. -
Spectrometer optical system structure needs to consider the working wavelength range, spectral resolution, luminous flux, volume and its match with the detector and other factors[17], the main performance parameters of the designed spectrometer are shown in Tab.1.
Table 1. Main parameters of a DMD-based small NIR spectrometer
Parameter Value Wavelength range 1000-1700 nm Spectral resolution <15 nm Size <150 mm×150 mm×150 mm -
In order to realize the miniaturization of the spectroscopy instrument and satisfy the requirement that multiple channels of data can be collected at one time, a receiver system with high sensitivity and low dark noise should be selected. The InGaAs detector manufactured by Pioneer Technology Inc. is selected as the spectral receiver which has a PDA that is sensitive to near-infrared light and meets the needs of the band of 1000-1700 nm. Its spectral response curve is shown in Fig.2.
-
The DMD-based small NIR spectrometer system uses a digital micromirror manufactured by Texas Instruments. The DLP4500NIR model has the advantages of fast switching speed and high reliability. It consists of 1140×912 coated digital micromirrors, each with a size of 7.6 μm×7.6 μm and a diagonal dimension of 10.8 micrometer. Each micromirror can rotate ±12° around the axis of symmetry under the action of an electrostatic drive circuit. Each angle corresponds to a different working state. Its working principle is shown in Fig.3. It keeps the direction of incident light unchanged. When the direction angle between incident light and the normal direction of the DMD surface is 24°, the light of the DMD surface is also emitting at 24° on the other side of the normal, rather than entering the optical system. When the incident light is +12° to the normal direction of the DMD, that is, the micromirror plane is rotated by 12°, the incident light is reflected by the DMD surface and enters the subsequent optical system; when the incident light and the DMD surface normal direction are 36°, the plane of the micromirror is rotated by −12°, and the reflected light and the incident light are emitted at 72°. Reflected light finally is absorbed by the light blocking plate.
-
The grating is a very important element in the spectrometer and can directly affect the resolution of the system. We select a planar reflective grating for spectral dispersion, from the grating equation:
$$\begin{array}{l} {\rm{d}}(\sin \theta \pm \sin i) = m\lambda \\ (m = 0, \pm 1, \pm 2,\cdots) \\ \end{array} $$ (7) where d is the grating constant, i and θ are the incident angle and the diffraction angle, respectively, and m is the diffraction order of the spectrum[11]. Angular dispersion rate of grating can be expressed as:
$$\frac{{{\rm{d}}\theta }}{{{\rm{d}}\lambda }}=\frac{m}{{{\rm{d}}\cos \theta }}$$ (8) The dispersion rate of grating line is
$$\frac{{{\rm{d}}l}}{{{\rm{d}}\lambda }}=\frac{m}{{{\rm{d}}\cos \theta }}f'$$ (9) where
$ {f}' $ represents the focal length of the imaging lens behind the grating. In this design, the line dispersion rate and the size of the DMD's micro-mirror determine the spectral resolution of the spectrometer. According to the theory of diffraction grating, m=1, d=300 l/mm, the blaze angle is 13.0°, and the grating diffraction efficiency curve of the aluminized film is shown in Fig.4. -
According to the Eqs. (1)-(9), a theoretical calculation model is established by combining the parameters of the detector, the grating, and the DMD. The structural parameters of the optical system obtained using the model are shown in Tab.2.
Table 2. Optical system design parameters
Parameter Value Working wavelength 1000-1700 nm Slit length 1 mm Slit width 50 μm Collimating lens focal length 75 nm Imaging lens focal length 45 nm Raster line density 300 l/mm DMD pixel size 7.6 μm×7.6 μm DMD pixel number 1140×912 pixels DMD tilt angle ±12°
Principle and optimum analysis of small near-infrared spectrometers based on digital micromirror device
-
摘要: 数字微镜器件小型近红外光谱仪器具有检测速度快、灵敏度高、无损伤检测、仪器小型化等优点,广泛应用于化学成分分析和质量检测。然而,作为仪器设计的前提,整个光谱范围的光学优化设计是系统的难点。文中研究了基于数字微镜器件小型近红外光谱仪的光谱成像系统的理论设计方法。该方法采用双配消像差透镜,结合几何像差理论,对小型数字微镜器件近红外光谱仪的设计进行了优化,以降低整个系统的像差。然后,利用光学仿真软件对准直接成像系统进行光学仿真。最终达到设计仿真要求。仿真结果表明,该分光计的整体尺寸小于150 mm×150 mm×150 mm,在工作波段1000~1700 nm范围内,其分辨率优于15 nm。因此,该方法能满足设计要求,在实际应用中具有广阔的应用前景。Abstract: The DMD small near-infrared spectroscopy instrument is widely used in chemical composition analysis and quality inspection for its advantages of fast detection speed, high sensitivity, no damage detection, and miniaturization of portable instruments. However, as the premise of instrument design, optical optimization design of the whole spectral range is the hard work of the system. In this paper, the theoretical design method of the spectroscopic imaging system based on the small near-infrared spectrometer of DMD was studied. The method was designed by using the double-dispensing anti-aberration lens and combining the geometric aberration theory to optimize the design of a small DMD near-infrared spectrometer to reduce the aberration of the entire system. Then, the optical simulation software was used to align the direct imaging system for optical simulation. And ultimately achieve the design simulation requirements. Simulation results indicate that the whole size of the spectrometer is less than 150 mm×150 mm×150 mm, and the resolution is better than 15 nm in the range of 1000-1700 nm in the working band. Therefore, the proposed method can meet the design requirements and has broad application prospects in practical applications.
-
Table 1. Main parameters of a DMD-based small NIR spectrometer
Parameter Value Wavelength range 1000-1700 nm Spectral resolution <15 nm Size <150 mm×150 mm×150 mm Table 2. Optical system design parameters
Parameter Value Working wavelength 1000-1700 nm Slit length 1 mm Slit width 50 μm Collimating lens focal length 75 nm Imaging lens focal length 45 nm Raster line density 300 l/mm DMD pixel size 7.6 μm×7.6 μm DMD pixel number 1140×912 pixels DMD tilt angle ±12° -
[1] Xue Qingsheng, Wang Shurong, Li Futian. High precision linearity measurement instrument for ultraviolet photodetectors [J]. Optics and Precision Engineering, 2008, 35(7): 116-120. (in Chinese) [2] Wang Xin, Ding Xuezhuan, Yang Bo, et al. Optical system design and spectral characteristics calculation of prism spectrometer [J]. Acta Optica Sinica, 2010, 39(7): 1334-1339. (in Chinese) [3] Jin Zhao, Xiao Peng, Dai Jingmin. Development of a broadband infrared grating scan spectral radiometer [J]. Infrared Technology, 2005, 27(4): 343-346. (in Chinese) [4] Han Tao, Ni Weiming, Chen Liangyao. New pattern monochromator of multi-grating using two-dimensional color CCD array [J]. Waves, 2002, 21(S): 118-122. (in Chinese) [5] Gong Yan, Chen Bo. Optical design of an accurate focusing grazing incidence spherical grating monochromator [J]. Acta Optica Sinica, 2001, 21(5): 574-577. (in Chinese) doi: 10.3321/j.issn:0253-2239.2001.05.016 [6] Wu Guoan. Design of Spectrometry[M]. Beijing: Science Press, 1978. (in Chinese) [7] Li Quanchen, Jiang Yuejuan. The Theory of Spectrometry[M]. Beijing: Publishing House of Beijing Institute of Technology, 1999. (in Chinese) [8] Chu Xiaoli, Tian Songbai, Xu Yupeng, et al. Study on rapid evaluation of crude oil by near infrared spectroscopy[J]. Petroleum Processing and Petrochemicals , 2012, 43(1): 89-93. doi: 10.3969/j.issn.1005-2399.2012.01.019 [9] Herschel W. Investigation of the power of the prismatic colours to heat and illuminate objects [J]. Philosophical Transactions of the Royal Society, 1880, 90: 255-258. [10] O'Brien N A, Hulse C A, Friedrich D M. Miniature near-infrared (NIR)spectrometer engine for handheld applications[C]//SPIE, 2012, 8374(7): 3. [11] [12] Fan Shifu. Recent development of spectral technology and instruments[J]. Modern Scientific Instruments , 2006(5): 14-19. doi: 10.3969/j.issn.1003-8892.2006.05.004 [13] Wang Xingquan, Feng Kecheng, Zhu Guoxian, et al. Study on the calculation formula of resolution and wavelength range of grating spectrometer [J]. Acta Optica Sinica, 2009, 38(5): 1167-1170. (in Chinese) [14] Xu Jialin, sun Qiang. Research on key technologies of Hadamard transform near infrared spectrometer based on DMD[D]. Beijing: Graduate School of Chinese Academy of Sciences, 2017. (in Chinese) [15] [16] [17] Tian Gaoyou, Chu Xiaoli, Yuan Hongfu, et al. Summary of main technical indexes and evaluation methods of near infrared spectrometer[J]. Modern Scientific Instruments , 2005(4): 17-20. doi: 10.3969/j.issn.1003-8892.2005.04.007