Volume 46 Issue 1
Feb.  2017
Turn off MathJax
Article Contents

Liu Zhen, Hong Jin, Gong Guanyuan, Zheng Xiaobing, Yang Weifeng, Yuan Yinlin. Alignment error optimization of spatially modulated imaging polarimeter system[J]. Infrared and Laser Engineering, 2017, 46(1): 117003-0117003(7). doi: 10.3788/IRLA201746.0117003
Citation: Liu Zhen, Hong Jin, Gong Guanyuan, Zheng Xiaobing, Yang Weifeng, Yuan Yinlin. Alignment error optimization of spatially modulated imaging polarimeter system[J]. Infrared and Laser Engineering, 2017, 46(1): 117003-0117003(7). doi: 10.3788/IRLA201746.0117003

Alignment error optimization of spatially modulated imaging polarimeter system

doi: 10.3788/IRLA201746.0117003
  • Received Date: 2016-05-10
  • Rev Recd Date: 2016-06-20
  • Publish Date: 2017-01-25
  • Spatially modulated imaging polarimeter (SMIP) system is able to modulate four Stokes parameters S0-S3 in a single interferogram by using Savart plate. Through sliding reconstruction method, the entire polarization information can be demodulated from the interference fringes. In this system, the alignment errors of the half wave plate (HWP) and the analyzer has a non-ignorable impact on the measurement accuracy of the Stokes parameters. A theoretical model including these two alignment errors were presented in this paper. Based on this model and parameters of the prototype SMIP, the Stokes parameters measurement accuracy of unpolarized light, 0/90linear polarized light, 45/135linear polarized light and left/right circularly polarized light was analyzed. According to these four basic polarized lights, a method for solving Stokes parameters measurement error of incident light with any state of polarization (SOP) and degree of polarization (DOP) was given. In order to alleviate the influence of HWP and analyzer alignment errors, an optimization using condition number of measurement matrix as an objective function was given. It is shown that when the thickness of Savart plate is 23 mm, which is corresponding to the minimum condition 2.06, measurement errors induced by HWP and analyzer alignment errors can be effectively diminished.
  • [1] Sun X, Qiao Y, Hong J. Review of polarization remote sensing techniques and applications in the visible and infrared[J]. Journal of Atmospheric and Environmental Optics, 2010, 3:5.
    [2] Li Yanan, Sun Xiaobing, Mao Yongna, et al. Spectral polarization characteristic of space target[J]. Infrared and Laser Engineering, 2012, 41(1):205-210. (in Chinese).
    [3] [3 Tyo J S, Goldstein D L, Chenault D B, et al. Review of passive imaging polarimetry for remote sensing applications[J]. Applied optics, 2006, 45(22):5453-5469.
    [4] Oka K, Kaneko T. Compact complete imaging polarimeter using birefringent wedge prisms[J]. Optics Express, 2003, 11(13):1510-1519.
    [5] Oka K, Saito N. Snapshot complete imaging polarimeter using Savart plates[C]//SPIE Optics Photonics, 2006, 6295:629508.
    [6] Oka K, Suda R, Ohnuki M, et al. Snapshot imaging polarimeter for polychromatic light using Savart plates and diffractive lenses[J]. Frontiers in Optics, 2009:FThF4.
    [7] Cao Q, Zhang C, DeHoog E. Snapshot imaging polarimeter using modified Savart polariscopes[J]. Appl Opt, 2012, 51(24):5791-5796.
    [8] Saito N, Odate S, Otaki K, et al. Wide field snapshot imaging polarimeter using modified Savart plates[C]//SPIE Optical Engineering Applications. International Society for Optics and Photonics, 2013:88730M.
    [9] Hanchen W J L. Influence of transmissivity in Savart polariscope on throughput of spatially modulated polarization interference imaging spectroscope[J]. Acta Optica Sinica, 2009, 1:17.
    [10] Zhang ChunMin, Ren Wenyi, Mu Tingkui. Effects of optical axis direction on optical path difference and lateral displacement of Savart polariscope[J]. Chinese Physics B, 2010, 19(2):024202. (in Chinese)
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(450) PDF downloads(148) Cited by()

Related
Proportional views

Alignment error optimization of spatially modulated imaging polarimeter system

doi: 10.3788/IRLA201746.0117003
  • 1. Anhui Institute of Optics and Fine Mechanics,Chinese Academy of Sciences,Hefei 230031,China;
  • 2. University of Science and Technology of China,Hefei 230026,China;
  • 3. Key Laboratory of Optical Calibration and Characterization,Chinese Academy of Sciences,Hefei 230031,China;
  • 4. University of Chinese Academy of Sciences,Beijing 100049,China

Abstract: Spatially modulated imaging polarimeter (SMIP) system is able to modulate four Stokes parameters S0-S3 in a single interferogram by using Savart plate. Through sliding reconstruction method, the entire polarization information can be demodulated from the interference fringes. In this system, the alignment errors of the half wave plate (HWP) and the analyzer has a non-ignorable impact on the measurement accuracy of the Stokes parameters. A theoretical model including these two alignment errors were presented in this paper. Based on this model and parameters of the prototype SMIP, the Stokes parameters measurement accuracy of unpolarized light, 0/90linear polarized light, 45/135linear polarized light and left/right circularly polarized light was analyzed. According to these four basic polarized lights, a method for solving Stokes parameters measurement error of incident light with any state of polarization (SOP) and degree of polarization (DOP) was given. In order to alleviate the influence of HWP and analyzer alignment errors, an optimization using condition number of measurement matrix as an objective function was given. It is shown that when the thickness of Savart plate is 23 mm, which is corresponding to the minimum condition 2.06, measurement errors induced by HWP and analyzer alignment errors can be effectively diminished.