Li Ming, Zong Xiaoying. In-lab system-level BRDF measurement method of calibration diffuser[J]. Infrared and Laser Engineering, 2017, 46(1): 117004-0117004(7). doi: 10.3788/IRLA201746.0117004
Citation: Li Ming, Zong Xiaoying. In-lab system-level BRDF measurement method of calibration diffuser[J]. Infrared and Laser Engineering, 2017, 46(1): 117004-0117004(7). doi: 10.3788/IRLA201746.0117004

In-lab system-level BRDF measurement method of calibration diffuser

doi: 10.3788/IRLA201746.0117004
  • Received Date: 2016-05-10
  • Rev Recd Date: 2016-06-20
  • Publish Date: 2017-01-25
  • The on-board calibration unit is an important part of the space-borne remote sensor and used to achieve the on-board radiometric calibration and other functions. An on-board calibration unit installed on a time-modulated fourier transform spectrometer was introduced. Used the sun as the light source to illuminate the calibration diffuser, the on-board calibration unit carries out the on-board radiometric calibration of the spectrometer in the full aperture, the full optical path and the full field of view. The calibration diffuser was in the front of the optical path of the spectrometer to reflect the sunlight and the standard radiance was obtained by used the known exoatmospheric solar irradiance and the BRDF(bidirectional reflectance distribution function) of the calibration diffuser. The BRDF of the calibration diffuser needs to be measured accurately in the lab. With the whole spectrometer, using the sun simulator and the standard diffuser, the in-lab system-level BRDF test method of the calibration diffuser was introduced. While using the system-level BRDF of the calibration diffuser, the accuracy of the on-board absolute radiometric calibration was analyzed and can satisfy the requirement of 5%.
  • [1] Han Qijin, Zhang Xuewen, Qiao Zhiyuan, et al. Wide dynamic radiometric calibration of GF-1 PMS sensors using multi-test sites[J]. Infrared and Laser Engineering, 2015, 44(1):127-133. (in Chinese)韩启金, 张学文, 乔志远, 等. 高分一号卫星PMS相机多场地宽动态辐射定标[J]. 红外与激光工程, 2015, 44(1):127-133.
    [2] Han Qijin, Fu Qiaoyan, Pan Zhiqiang, et al. Absolute radiometric calibration and validation analysis of ZY-3 using artificial targets[J]. Infrared and Laser Engineering, 2013, 42(S1):167-173. (in Chinese)韩启金, 傅俏燕, 潘志强, 等. 资源三号卫星靶标法绝对辐射定标与验证分析[J]. 红外与激光工程, 2013, 42(S1):167-173.
    [3] Xiaoxiong X, Jim B. MODIS solar reflective calibration traceability[C]//SPIE, 2009, 7452:74520K.
    [4] Yang Benyong, Zhang Liming, Shen Zhengguo, et al. Characteristic measurement of diffuser panel used for on-board calibration for satellite optical sensors[J]. Optics and Precision Engineering, 2009, 17(8):1851-1858. (in Chinese)杨本永, 张黎明, 沈政国, 等. 光学传感器星上定标漫射板的特性测量[J]. 光学精密工程, 2009, 17(8):1851-1858.
    [5] Wang Long, Lin Chao, Zheng Yuquan. Fabrication and experiment of aluminum diffuser for CO2 detector calibration on orbit[J]. Chinese Optics, 2013, 6(4):591-599. (in Chinese)王龙,蔺超,郑玉权. CO2探测仪星上定标铝漫反射板的制备与试验[J]. 中国光学, 2013, 6(4):591-599.
    [6] Robert A B, Steven W B, Keith R L, et al. System-level pre-launch calibration of on-board solar diffusers[C]//SPIE, 2007, 6677:667704.
    [7] Marcel R D, Ruud J D, Pieternel F L, et al. Ozone monitoring instrument calibration[J]. IEEE Transactions on Geosciences and Remote Sensing, 2006, 44(5):1209-1238.
    [8] Wang Min, He Mingyuan, Zhang Shuiping, et al. Calibration methods of visible-near infrared channel-type satellite radiometric sensor[J]. Remote Sensing Information, 2014, 29(1):114-120. (in Chinese)王敏, 何明元, 张水平, 等. 通道式可见光近红外卫星遥感器辐射定标方法综述[J]. 遥感信息, 2014, 29(1):114-120.
    [9] Zhang Baogui, Zhang Yufeng. Analysis of relative radiometric calibration algorithm of space CCD camera[J]. Spacecraft Recovery Remote Sensing, 2014, 35(1):91-96. (in Chinese)张保贵, 张宇烽. 空间CCD相机辐射校正算法分析[J]. 航天返回与遥感, 2014, 35(1):91-96.
    [10] Li Ning, Zhang Yunfeng, Liu Chunxiang, et al. Calibration of 1 m aperture infrared theodolite[J]. Optics and Precision Engineering, 2014, 22(8):2054-2060. (in Chinese)李宁, 张云峰, 刘春香, 等. 1 m口径红外测量系统的辐射定标[J]. 光学精密工程, 2014, 22(8):2054-2060.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(675) PDF downloads(160) Cited by()

Related
Proportional views

In-lab system-level BRDF measurement method of calibration diffuser

doi: 10.3788/IRLA201746.0117004
  • 1. Beijing Institute of Space Mechanics & Electricity,Beijing 100190,China

Abstract: The on-board calibration unit is an important part of the space-borne remote sensor and used to achieve the on-board radiometric calibration and other functions. An on-board calibration unit installed on a time-modulated fourier transform spectrometer was introduced. Used the sun as the light source to illuminate the calibration diffuser, the on-board calibration unit carries out the on-board radiometric calibration of the spectrometer in the full aperture, the full optical path and the full field of view. The calibration diffuser was in the front of the optical path of the spectrometer to reflect the sunlight and the standard radiance was obtained by used the known exoatmospheric solar irradiance and the BRDF(bidirectional reflectance distribution function) of the calibration diffuser. The BRDF of the calibration diffuser needs to be measured accurately in the lab. With the whole spectrometer, using the sun simulator and the standard diffuser, the in-lab system-level BRDF test method of the calibration diffuser was introduced. While using the system-level BRDF of the calibration diffuser, the accuracy of the on-board absolute radiometric calibration was analyzed and can satisfy the requirement of 5%.