Volume 47 Issue 6
Jul.  2018
Turn off MathJax
Article Contents

Ho Bowei, Liang Yizhi, Hsu Chelun, Das Subir, Kao Fujen. Stimulated gain and spontaneous loss pump-probe microscopy[J]. Infrared and Laser Engineering, 2018, 47(6): 606004-0606004(6). doi: 10.3788/IRLA201847.0606004
Citation: Ho Bowei, Liang Yizhi, Hsu Chelun, Das Subir, Kao Fujen. Stimulated gain and spontaneous loss pump-probe microscopy[J]. Infrared and Laser Engineering, 2018, 47(6): 606004-0606004(6). doi: 10.3788/IRLA201847.0606004

Stimulated gain and spontaneous loss pump-probe microscopy

doi: 10.3788/IRLA201847.0606004
  • Received Date: 2018-01-05
  • Rev Recd Date: 2018-02-03
  • Publish Date: 2018-06-25
  • In this work, the pump-probe microscope was used to study the stimulated gain and spontaneous loss phenomenon. A pulsed diode laser, pu=635 nm as the pump (excitation) beam and a mode-locked Ti-sapphire laser, pr=780 nm, as the probe (stimulation) beam were applied. For stimulated gain, the pump beam was modulated at a frequency, f1, and the probe beam was demodulated accordingly to extract the signal in the transmission direction with a photodiode as the detector (PDA 36A, Thorlabs). For spontaneous loss, the probe beam was modulated at frequency, f2, the spontaneous loss signal was then demodulated from the fluorescence detected in the reflection mode by a PMT. In all cases, a high performance lock-in amplifier (HF2LI, Zurich Instruments) was used. The output signal of the lock-in amplifier was then fed to the A/D channel of the scanning unit for image reconstruction. The scan rate was set at a frequency 500 Hz, to match the time constant (1.99 ms) of the lock-in amplifier. By demodulating fluorescence signal, the fluorescence lifetime and optical section images can be obtained with greatly reduced background, in which shot noise was attributed. Additionally, the signal-to-noise ratio was improve and penetration depth like multiphoton microscopy was enhanced, without expansive femtosecond lasers.
  • [1] Fischer M C, Wilson J W, Robles F E, et al. Invited Review Article:Pump-probe microscopy[J]. Review of Scientific Instruments, 2016, 87(3):031101.
    [2] Tian Peifang, Warren W S. Ultrafast measurement of two-photon absorption by loss modulation[J]. Optics Letters, 2002, 27(18):1634-1636.
    [3] Fu Dan, Tong Ye, Matthews T E, et al. High-resolution in vivo imaging of blood vessels without labeling[J]. Optics Letters, 2007, 32(18):2641-2643.
    [4] Matthews T E, Matthews I R, Selim M A, et al. Pump-probe imaging differentiates melanoma from melanocytic nevi[J]. Science Translational Medicine, 2011, 3(71):71ra15.
    [5] Lin Poyen, Lee Shinshian, Chang Chiaseng, et al. Long working distance fluorescence lifetime imaging with stimulated emission and electronic time delay[J]. Optics Express, 2012, 20(10):11445-11450.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(486) PDF downloads(46) Cited by()

Related
Proportional views

Stimulated gain and spontaneous loss pump-probe microscopy

doi: 10.3788/IRLA201847.0606004
  • 1. Institute of Biophotonics,College of Biomedical and Engineering,National Yang-Ming University,Taipei 11221,China

Abstract: In this work, the pump-probe microscope was used to study the stimulated gain and spontaneous loss phenomenon. A pulsed diode laser, pu=635 nm as the pump (excitation) beam and a mode-locked Ti-sapphire laser, pr=780 nm, as the probe (stimulation) beam were applied. For stimulated gain, the pump beam was modulated at a frequency, f1, and the probe beam was demodulated accordingly to extract the signal in the transmission direction with a photodiode as the detector (PDA 36A, Thorlabs). For spontaneous loss, the probe beam was modulated at frequency, f2, the spontaneous loss signal was then demodulated from the fluorescence detected in the reflection mode by a PMT. In all cases, a high performance lock-in amplifier (HF2LI, Zurich Instruments) was used. The output signal of the lock-in amplifier was then fed to the A/D channel of the scanning unit for image reconstruction. The scan rate was set at a frequency 500 Hz, to match the time constant (1.99 ms) of the lock-in amplifier. By demodulating fluorescence signal, the fluorescence lifetime and optical section images can be obtained with greatly reduced background, in which shot noise was attributed. Additionally, the signal-to-noise ratio was improve and penetration depth like multiphoton microscopy was enhanced, without expansive femtosecond lasers.

Reference (5)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return