  • Publish Date: 2019-05-25
  • Atom Trap Trace Analysis(ATTA) technology, which is based on the theory of laser cooling and trapping, has the capability of high-sensitivity detection of the radioactive isotope of Krypton and wide applications in the fields of geophysics and environmental science. Zeeman slower, as a key component of the ATTA instrument, is used to generate continuous atomic beam with low velocity. With the advantages of stable magnetic distribution, easy installment and debugging, no constant current power or cooling requirement, the Zeeman slower based on permanent magnet is getting more and more attention in recent years. In this paper, a Zeeman slower based on the toroidal permanent magnet was designed, the spatial distribution of the magnetic field of this slower was calculated by finite element analysis, a prototype was manufactured according to the design parameters, and its magnetic field along the axis was also measured. The lengths of the slower and its effective deceleration area were 51.2 cm and 46.9 cm, separately. The maximum deviation between the measured and theoretical magnetic field was less than 3.6 G, and the average deviation was 1.3 G. Furthermore, the deceleration process of the atomic beam in the designed and actual magnetic field distribution was simulated, and the influence of the radial variation of magnetic field distribution on the deceleration process of the atomic beam was analyzed. The result shows that the Zeeman slower in this paper is able to decelerate the velocity of the atomic beam with a diameter less than 20 mm from the maximum initial value of 250 m/s to the final value of 50 m/s.
  • [1] 王义遒. 原子的激光冷却与陷俘[M]. 北京:北京大学出版社, 2007.
    [2] Zhang Yan, Zhang Xiaohang, Zhang Yu, et al. Generation and modulation of high efficiency stationary optical signals in cold 87Rb atomic samples[J]. Chinese Optics, 2012, 5(2):143-147. (in Chinese)张岩, 张晓航, 张宇, 等. 冷铷原子样品中高效静态光信号的生成与调制[J]. 中国光学, 2012, 5(2):143-147.
    [3] Fan Pengge, Wu Yiming, Jia Sen, et al. Optimization design of two dimensional magneto optical trap field[J]. Infrared and Laser Engineering, 2016, 45(6):0618003. (in Chinese)樊鹏格, 吴易明, 贾森, 等. 冷原子干涉仪中二维磁光阱线圈的优化设计[J]. 红外与激光工程, 2016, 45(6):0618003.
    [4] Guo Xiuzhen, Hou Lixin, Yin Zhaotai, et al. All-optical routing control based on coherently induced high reflection band and high trensmission band in a medium of cold atoms[J]. Chinese Optics, 2011, 4(4):355-362. (in Chinese)国秀珍, 侯丽新, 尹昭泰, 等. 冷铷原子介质中基于相干诱导高反射带和高透射带的全光路由控制[J]. 中国光学, 2011, 4(4):355-362.
    [5] Ren Jie, Liu Hui, Lu Benquan, et al. Program control in transition observation of Strontium optical lattice clock[J]. Optics and Precision Engineering, 2016, 24(1):50-58. (in Chinese)任洁, 刘辉, 卢本全, 等. 锶原子光钟钟跃迁谱线探测中的程序控制[J]. 光学精密工程, 2016, 24(1):50-58.
    [6] Yang Guomin, Tong Amin, Dong Xize, et al. Dating and tracing groundwater and ice with 81Kr and 85Kr[C]//19th EGU General Assembly, EGU2017, 2017, 11467:23-28.
    [7] Chen C Y, Li Y M, Bailey K, et al. Ultrasensitive isotope trace analyses with a Magneto-Optical trap[J]. Science, 1999, 286(5442):1139-1141.
    [8] Li Jie, Pang Zhonghe, Yang Guomin, et al. Million-year-old groundwater revealed by krypton-81 dating in Guanzhong Basin, China[J]. Science Bulletin, 2017(17):1181-1184.
    [9] Lu Z T, Schlosser P, Jr W S, et al. Tracer applications of noble gas radionuclides in the geosciences[J]. Earth-Science Reviews, 2014, 138:196-214.
    [10] Metcalf H J, Peter Van Der Straten. Laser Cooling and Trapping[M]. Switzerland:Springer, 1999.
    [11] Lu Xuanhui, Wang Jiangfeng. Quantum gyroscope based on an atom interferometer[J]. Infrared and Laser Engineering, 2007, 36(3):293-295. (in Chinese)陆璇辉, 王将峰. 基于原子干涉的量子陀螺仪[J]. 红外与激光工程, 2007, 36(3):293-295.
    [12] Wang Zhanshan, Ma Shanshan, Ma Yan, et al. Knife-edge technique for laser cooling[J]. Optics and Precision Engineering, 2006, 14(1):63-69. (in Chinese)王占山, 马珊珊, 马艳, 等. 刀口技术在激光冷却中的应用[J]. 光学精密工程, 2006, 14(1):63-69.
    [13] Hill I R, Ovchinnikov Y B, Elizabeth M B, et al. A simple, configurable, permanent magnet Zeeman Slower for Sr[C]//Conference:European Frequency and Time Forum (EFTF),2012:545-549.
    [14] Hill I R, Ovchinnikov Y B, Elizabeth M B, et al. Zeeman slowers for Strontium based on permanent magnets[J]. Physics, 2014, 47(7):1216-1221.
    [15] Zhang Xiaohang, Xu Xinye. Development of adjustable permanent magnet Zeeman slowers for optical lattice clocks[J]. Chinese Physics B, 2017, 26(5):102-107.
    [16] Zhang Xiaohang, Xu Xinye. Optimized design of a permanent Zeeman slower for an Ytterbium optical lattice clock[J]. Laser Physics, 2016, 26(7):075501.
    [17] Krzyzewski S P, Akin T G, Dahal P, et al. A clip-on zeeman slower using toroidal permanent magnets[J]. Review of Scientific Instruments, 2014, 85(10):103104.
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(556) PDF downloads(31) Cited by()

Related
Proportional views

A toroidal permanent magnet Zeeman slower for Krypton

doi: 10.3788/IRLA201948.0520004
  • 1. State Key Laboratory of Laser Propulsion & Application,Space Engineering University,Beijing 101416,China

Abstract: Atom Trap Trace Analysis(ATTA) technology, which is based on the theory of laser cooling and trapping, has the capability of high-sensitivity detection of the radioactive isotope of Krypton and wide applications in the fields of geophysics and environmental science. Zeeman slower, as a key component of the ATTA instrument, is used to generate continuous atomic beam with low velocity. With the advantages of stable magnetic distribution, easy installment and debugging, no constant current power or cooling requirement, the Zeeman slower based on permanent magnet is getting more and more attention in recent years. In this paper, a Zeeman slower based on the toroidal permanent magnet was designed, the spatial distribution of the magnetic field of this slower was calculated by finite element analysis, a prototype was manufactured according to the design parameters, and its magnetic field along the axis was also measured. The lengths of the slower and its effective deceleration area were 51.2 cm and 46.9 cm, separately. The maximum deviation between the measured and theoretical magnetic field was less than 3.6 G, and the average deviation was 1.3 G. Furthermore, the deceleration process of the atomic beam in the designed and actual magnetic field distribution was simulated, and the influence of the radial variation of magnetic field distribution on the deceleration process of the atomic beam was analyzed. The result shows that the Zeeman slower in this paper is able to decelerate the velocity of the atomic beam with a diameter less than 20 mm from the maximum initial value of 250 m/s to the final value of 50 m/s.

Reference (17)

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return