Volume 42 Issue 4
Feb.  2014
Turn off MathJax
Article Contents

Chen Yinghang, Sui Zuoning, Sun Hongsheng, Wei Jianqiang, Li jingfeng, Li Shiwei, Sun Guangwei. Facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance[J]. Infrared and Laser Engineering, 2013, 42(4): 998-1002.
Citation: Chen Yinghang, Sui Zuoning, Sun Hongsheng, Wei Jianqiang, Li jingfeng, Li Shiwei, Sun Guangwei. Facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance[J]. Infrared and Laser Engineering, 2013, 42(4): 998-1002.

Facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance

  • Received Date: 2012-08-01
  • Rev Recd Date: 2012-09-12
  • Publish Date: 2013-04-25
  • In order to adapt the fast development of ultraviolet detection and fitting the requirements for the calibration of ultraviolet optical sources and ultraviolet detectors, a facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance was set up. The facility was composed of standard optical source (deuterium lamp), standard detector(intensive silicon diode), ultraviolet monochromator, vacuum chamber and so on. The double standards and compared method were adopted in the facility to calibrate the equipments. As well as the value of the standard detector and source were transferred to the detector and the source which were calibrated. Spectral range is 110-400 nm.The maximum uncertainty for calibrating the detector is 12%(k=2), and for the optical source is 20%(k=2).
  • [1] Shaw Pingshine, larason Thomas C, Rajeev Gupta, at al. The new ultraviolet spectral responsivity scale based on cryogenic radiometry at Synchrotron Ultraviolet Radiation Facility Ⅲ[J]. Review of Scientific Instruments, 2001, 72(5): 2242-2247.
    [2]
    [3] Li Zhigang, Shaw Pingshine, Uwe Arp, at al. Long-term monitoring of the ultraviolet irradiance scale at the facility for irradiance calibration using synchrotrons[J]. Metrologia, 2010, 47: 429-434.
    [4]
    [5] Boyer C N, Osterman S N, Thonnard S E. An EUV and FUV calibration facility for SSULI[C]//SPIE, 1994, 2282: 107-115.
    [6]
    [7] Kalmanson Phillip C, Charles Sarner, Stefan Thonnard. Facility for the calibration, construction, and environmental testing of instruments and detectors in the EUV and FUV[C]//SPIE, 2002, 4485: 303-315.
    [8]
    [9] Xiong Limin, Liu Jinyuan, Xue Fengyi, at al. Ultraviolet and vacuum ultraviolet spectral radiance standard with synchrotron radiation source[J]. Acta Optica Sinica, 2006, 26(4): 547-550. (in Chinese)
    [10]
    [11] Lin Jinyuan, Xue Fengyi, Liao Ningfang. A UV-VUV radiation calibration facility based on synchrotron radiation[J]. Transaction of Beijing Institute of Technology, 2007, 27(9): 807-810. (in Chinese)
    [12]
    [13] Zhan Chunlian, Li Zhengqi, Liu Jianping, et al. Measurement of spectral irradiance[J]. Acta Photonica Sinica, 2009, 38(5): 1245-1249. (in Chinese)
  • 加载中
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Article Metrics

Article views(402) PDF downloads(258) Cited by()

Related
Proportional views

Facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance

  • 1. Beijing Zhenxing Metrology and Measurement Institute,Beijing 100074,China

Abstract: In order to adapt the fast development of ultraviolet detection and fitting the requirements for the calibration of ultraviolet optical sources and ultraviolet detectors, a facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance was set up. The facility was composed of standard optical source (deuterium lamp), standard detector(intensive silicon diode), ultraviolet monochromator, vacuum chamber and so on. The double standards and compared method were adopted in the facility to calibrate the equipments. As well as the value of the standard detector and source were transferred to the detector and the source which were calibrated. Spectral range is 110-400 nm.The maximum uncertainty for calibrating the detector is 12%(k=2), and for the optical source is 20%(k=2).